6.1.3

Potential Malfunctions

Trouble
"Screeching" noises
heard during the self-
test or during use of
®
the SonicWeld Rx
system.
Sonotrode tip will not
detach from SonicPin.
SonicPin fails to attach
to sonotrode tip.
Sonotrode tip sticks to
SonicPin.
Handpiece indicator
does not light up
during activation.
Display dark.
66
Cause
Residual moisture or
condensate on the
contact points.
Bent or rammed
sonotrode tip.
Ultrasound applied
too long.
Bent sonotrode tip.
LED life exceeded.
No power supply.
Mains fuse defective.
Unit not switched on.
Instructions for Use / Technical Description
Identification
Clearly audible.
Visual check.
Pin is generally
deformed, sonotrode
cannot be removed
residue-free.
Visual check.
Visual check.
Visual check.
Electrical test.
Visual check.
Ultrasonic Unit SonicWeld Rx
Remedy
This does not compromise the
proper functioning of the
®
SonicWeld Rx
system. Noise
will subside automatically
during normal use of the
device. It may be helpful to
use longer drying times after
sterilization.
Replace sonotrode tip.
Observe required cooling-
down time. Rotate the
sonotrode axially (twist) to
break the connection, see
section 4.4, page 55.
Replace sonotrode tip.
Replace the handpiece. A
defective indicator on the
handpiece has no adverse
effect on ultrasound
application.
Connect the system to the
power supply and operate the
On/Standby switch.
Replace mains fuse.
Press On/Standby switch until
display lights up.
®