The connecting cable of the handpiece is approximately 2.95 m long. If this length is not
sufficient, an additional handpiece with a 6-m connecting cable is available as well.
Screw the sonotrode manually in place on the handpiece and use the open-end
wrench (12, front fold-out page) to check it for secure attachment (torque: max. 0.3 Nm).
Note that the wrench is shipped non-sterile, so it needs to be sterilized prior to use.
NOTICE
Disconnection from the supply system!
®
The SonicWeld Rx
power supply.
The ultrasonic unit features 2 connecting sockets for handpieces (2, front fold-out page).
This enables you to connect 2 handpieces in order to work alternatingly with 2 sonotrodes
(e.g. 2 standard sonotrodes or 1 standard sonotrode and 1 smoothing sonotrode).
Plug the mains cable into the ultrasonic unit (5, front fold-out page) and then into a mains
socket-outlet with ground contact. As soon as the unit has been connected to the power
supply, it is automatically set to standby mode. Therefore, full disconnection is possible
only by pulling the plug of the mains cable out of the socket-outlet.
WARNING
Explosion hazard!
The SonicWeld Rx
mixtures of anesthetics and air, oxygen or nitrous oxide.
NOTICE
Make sure that no objects or appliances will be deposited on the ultrasonic unit.
Observe the recognized rules when connecting sterilized system components to non-
sterile equipment!
WARNING
Danger of injury due to malfunction!
®
SonicWeld Rx
must not be placed close to or stacked with other devices (always maintain
separation distance!), see section 9 "Guidelines and Manufacturer's Declaration Regarding
Electromagnetic Compatibility (EMC)", page 71.
If operation in close proximity to or in a stack with other devices is unavoidable, the
®
SonicWeld Rx
must be monitored to verify its proper performance in such a setup.
52
must be set up so that it can be easily and reliably disconnected from the
®
is not suitable for use in atmospheres enriched with flammable
Instructions for Use / Technical Description
Ultrasonic Unit SonicWeld Rx
®
V. 1.0