Instructions for Use / Technical Description
Ultrasonic Unit SonicWeld Rx
6.1.2

Handpiece Statuses

Message displayed
x
V. 1.0
®
Cause
No handpiece connected.
Temperature values measured on
handpiece not plausible.
Max. temperature reached on
handpiece.
The handpiece cycle counter
pictogram appears during self-test.
(240 ≤cycle counter <250)
The flashing handpiece cycle counter
pictogram appears during self-test.
(cycle counter ≥250)
Handpiece pushbutton defective.
Remedy
Connect handpiece.
Replace handpiece.
Message disappears automatically as
soon as the temperature drops below
70°C (158°F). Alternative: Replace
handpiece.
Message disappears automatically
once the self-test has been
completed.
Following completion of the self-test,
the message must be acknowledged
with the On/Standby switch.
Replace handpiece.
65