6.2.5
Replacing the Handpiece with Connecting Cable
Set SonicWeld Rx
•
no longer be activated.
— The display turns dark upon entering standby mode.
Remove the plug of the handpiece connecting cable from the handpiece connecting socket
•
(2, front fold-out page).
Connect a new handpiece with connecting cable (7, 8, front fold-out page; see section 7
•
"Accessories & Spare Parts", page 69) to the SonicWeld Rx
connecting cable fully into the handpiece connecting socket (2, front fold-out page) and
rotate to lock it in place.
Operate the On/Standby switch to set the SonicWeld Rx
•
Check the unit for proper functioning.
•
WARNING
Danger of injury in the event of using non-approved accessories!
Be sure to use only the genuine parts available from the manufacturer as specified in
section 7 "Accessories & Spare Parts", page 69. Using third-party accessories or spare parts
can increase interference emission and reduce the interference immunity of the
®
SonicWeld Rx
.
68
®
to standby mode by pressing the On/Standby switch. The unit can now
Instructions for Use / Technical Description
Ultrasonic Unit SonicWeld Rx
®
system: insert the plug of the
®
to working mode.
®
V. 1.0