4.3.4

Visual Check

Check visually that all components used are free of defects.
— Damaged components must not be used. Replace!
NOTICE
The tip of the sonotrode must not show any signs of mechanical damage. It must not be bent
or rammed (typical damage after the handpiece has fallen down).
4.3.5

Automatic Functional Test

CAUTION
Danger of burns!
If the sonotrode is allowed to make skin contact during the functional test, this can cause
burns.
Avoid contact with skin, eyes, etc.!
Self-test: Upon turning on the unit with the On/Standby switch (4, front fold-out page), the
handpieces (if connected) are ready for a self-test. This is indicated for each handpiece by the
flashing activation LED (9, front fold-out page) and the following symbol flashing on the
corresponding side of the display:
The self-test for this handpiece will be started upon operating the hand activation
pushbutton (10, front fold-out page) for the first time. Hold down the button until the display
message changes from Testing to Cooling, the activation LED stops flashing and the acoustic
signal changes. Be sure to keep the tip of the sonotrode out of contact with objects during this
process. If the test is successful, the unit is automatically set to working mode.
Checkmark
displayed (checkmark right or left, depending on which handpiece has been
tested): Means "no malfunction detected".
Generator-related error messages are displayed in text form, while handpiece states are
indicated by pictogram on the respective side (depending on handpiece), see section 6.1
"Troubleshooting", page 64.