6
Maintenance
6.1
Troubleshooting
6.1.1
Displays
Message displayed
Fehler Stromversorgung
[Power supply error]
Temperaturmessung defekt
[Temperature meter defective]
Übertemperatur Generator
[Generator overtemperature]
Gerätefehler
[Internal fault]
Störung Ultraschall
[Ultrasound trouble]
1. Handstück entlasten!
[Relieve pressure on
handpiece!]
2. On/Standby drücken!
[Press On/Standby!]
Schweißzeit > 60s
[Welding time]
1. Sonotrode festziehen!
[Tighten sonotrode!]
2. On/Standby drücken!
[Press On/Standby!]
Systemfehler
[Internal fault]
64
Cause
Generator error or fault.
Generator problem (during
welding process).
System overload (sonotrode tip
stuck/jammed, excessive
pressure).
Maximum welding time of 60 s
exceeded.
Sonotrode seated too loosely.
Monitoring of all important
program modules.
Instructions for Use / Technical Description
Ultrasonic Unit SonicWeld Rx
Remedy
Disconnect the unit from the
power supply, and then
reconnect it.
If the problem persists, return
unit to manufacturer, see
section 1.1, page 45.
Acknowledge by pressing the
On/Standby switch.
If the problem persists,
disconnect the unit from the
supply.
Acknowledge by pressing the
On/Standby switch, then clear
the handpiece problem and
operate the activation
pushbutton again
After a cooling-down time of 5 s,
the unit is ready for use again.
Tighten sonotrode with wrench
(12, front fold-out page), and
acknowledge error.
Disconnect the unit from the
power supply, and then
reconnect it.
If the problem persists, return
unit to manufacturer, see
section 1.1, page 45.
®
V. 1.0