Herunterladen Inhalt Inhalt Diese Seite drucken

Adjustment Of The Counter Weight To The Intended Load - Sartorius Stedim Biotech CERTOMAT Tplus Bedienungsanleitung

Inkubations-schuttelschrank
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2.4.2 Adjustment of the Counter Weight to the Intended
Load
Upon delivery the CERTOMAT® Tplus is preadjusted for
a load of 10 kg. If the intended load is different from
this, the counter weight has to be readjusted
correspondingly. Thus the vibrations, which may occur
during operation, can be minimized.
1.
Weigh the completely equipped tray with all shaking
containers attached.
2.
Switch off the CERTOMAT® Tplus at the main switch.
3.
To adjust the counter weight with the variable tool
for the load open the front flap under the glass doors
downwards. The front flap is fixed with magnets.
The variable tool is located behind this flap.
4.
Turn the shaking table by hand to its rearmost
position.
5.
Block the drive of the shaking table by putting the
delivered hexagon socket screw key through the
boring of the shaking table (Fig. 9 and 10, pos. 4).
Carefully turn the shaking table until the drive has
totally blocked.
Fig. 9::
Adjustment of the counter weight
corresponding to the load
All manuals and user guides at all-guides.com
4
3
1
6.
Put the variable tool (1) through the opening under the
front flap in the housing and guide it on the axis of the
counter weight (2);
7.
Turn the counter weight with the variable tool in the
clockwise direction (for load <10 kg) or turn it out
anticlockwise (load >10kg).
The adjustment of the counter weight which is
necessary for a certain load is marked on the variable
tool.
8.
Remove the hexagon socket screw key and the variable
tool and close the front flap.
9.
Mount the loaded tray.
10. Switch on the CERTOMAT® Tplus.
The CERTOMAT® Tplus starts only if both doors are
closed.
11. Test for the silent operation at your intended shaking
speed.
12. If a re-adjustment is necessary, stop the shaker, switch
off the CERTOMAT® Tplus, remove the tray and repeat
steps 3 – 11.
1.
Variable tool
2.
Counter weight
3.
Axis
4.
Hexagon socket screw key
11