4.2 Preparing and Inflating
the CultiBag RM
4.2.1 Unpacking the
CultiBag RM
4.2.2 Securing the CultiBag RM
to the Bag Holder RM
24
Menu Operation
Overview page:
When you press the "Overview" button, it takes you to this page.
The rocker display is only used in conjunction with the BIOSTAT
basic for data entry. If the rocker is attached to the BIOSTAT
Tower touch screen can be used for operation. However, the control mode of
the bioreactor must be set to "DCU Tower" in the menu item
Menu" on page 34.
Remove the sterile single-use CultiBag RM bioreactor from the plastic protective film.
Caution!
Do not damage the sterile cultivation chamber when removing it.
Carefully open the outside packaging. Do not use a sharp instrument as this
could damage the CultiBag RM.
1. Open the guide rails on the bag holder by releasing the lever on both sides of the
holder as shown in the figure.
2. Now place one or two CultiBags RM in the bag holder and press the plastic bars on
both sides of the bags into the now open guide rails of the bag holder. If you are
only placing one CultiBag RM in the holder, place it on the left side of the holder
to ensure error-free temperature measurement. For information regarding the
number and size of CultiBags RM that can be attached to the rocker, please see
section
"1.5.1 Properties, Specifications" on page 15.
t
3. Now press the lever down again to secure the CultiBag(s) RM.
It displays the most
important parameters as
actual values. A green
LED next to the nominal
values indicates whether
or not the respective
controller has been
started.
Pressing the blue Home
button returns you to
the Main page.
RM 20 | 50
®
RM Tower, the
®
"4.4.7 Settings
t