12
Technical Data, Ordering Information
12.1
Technical Data
12.1.1 Mechanical Design
Shaking cabinet
Weight (with equipment)
Tray support
Materials
Trays, type/size
Load
12.1.2 Mains Supply
Laboratory connector
Mains specifications
Power consumption
Photo illumination
Working light
Fuses
Radio interfer.
suppression
12.1.3 Operating Data
Orbital motion
Shaking speed
Rotation speed accuracy
Incubation temperature
Accuracy of
incubation temperature
Ambient temperature
Ambient humidity
12.1.4 Analog Out
Socket, 15-poles Sub-D
Exactness
12.1.5 Collective Alarm
Collective alarm
12.1.6 Option Cooling Unit
Refrigerant
Filling
RT = room temperature
(We reserve the right to carry out dimensional and construction modifications)
1)
Trays of modified design for actual locking; rebuilding of old trays for this locking on request.
Please contact the Sartorius AG customer service.
All manuals and user guides at all-guides.com
outside housing 1265 x 710 x 775 mm (W x H x D);
incubation room 890 x 495 x 650 mm (W x H x D)
approx. 230 kg
440 x 420 mm (W x D)
outside cabinet : sheet steel case with acidproof varnish;
incubation room inside: stainless steel case
1)
Type E/EU (420 x 420 mm)
; type F/FU : (800 x 420 mm)
depending on tray and mounting system of the Sartorius Stedim Biotech GmbH
accessory program
laboratory grounded socket (specific connectors on request)
230 V (±5 %) 50 Hz or 115 V (±5 %) 60 Hz
about 1.035 kW (including refrigeration)
approx. 0.39 kW
approx. 0.0023 kW
T6,3A for 230 V, T10A for 115 V
according to EN 61326-1:2006
Ø 25 mm or Ø 50 mm, depending shaker drive
-1
40 ... 400 min
max. ±1% (of final value)
heating version
heating and cooling version
± 0.1 °C
+10 ... +40 °C (+10 ... + 30 °C at UHK version)
10 ... 60 %; Incubation room 10 ... 80 %, not condensing
0 ... 10 V, description see chapter 11.2.1
± 0,3 V
Contact without potential (directional contact), max. 230 VAC (6.5 A Ohm load)
R 134a
270g
1)
RT +8 °C ... +70 °C
10 °C under RT .. +70 °C
49