Select the Control Type

The correct control type must be selected before the control test is initiated.
From the Quality Control Menu:
1. To select the type of control to be performed, select [Control Type].
2. To view the complete list of choices, continue to press [Control Type].
3. To confirm the selection, press [Enter]. This selection will display only the
available control types for the test selected in the Main Menu.
Manually Enter a Control Lot Number, Expiration Date,
and Range
From the Quality Control Menu:
1. Select [Control Lot].
2. To select the desired control, highlight the control in the list.
When there are no lot numbers entered for a control, the following variable function
keys appear: [Add Lot Number], [Add Exp Date], [Set Range], and [Exit to
Quality Control Menu].
1. To enter the lot number, select [Add Lot Number].
2. Enter the lot number using the keypad.
3. To confirm the selection, press [Enter].
4. To enter the expiration date for the selected lot number, select [Add Exp
Date]. The date format is yy-mm-dd.
5. Enter the expiration date.
6. To confirm the entry, press [Enter].
7. To enter the range for the control, select [Set Range]. The format for entering
the control range is "lll-hhh," where "lll" is the low end and "hhh" is the high end
of the control range.
Note: If the low range is less than 100, a zero must precede the 2-digit
number, eg, "099."
8. Enter the range.
9. To confirm the entry, press [Enter].
10. To return to the "Add Lot Number" selection, select [Exit Add Selection].
When a control type has a single lot number entered and a second lot is to be
entered, the following variable function keys appear: [Add Lot/EXP Date],
[Remove Lot], [Edit Range] and [Exit to Quality Control Menu].
1. To add a second control lot for a cartridge type, select [Add Lot/EXP Date].
2. To enter the second lot number, see the instructions above (when there are
no lot numbers).
26 English
86702-001
Rev 1.0
Medtronic Confidential
ACT Plus™ Operator's Manual