Herunterladen Inhalt Inhalt Diese Seite drucken

General Quality Control (Qc) Functions - Medtronic ACT Plus Benutzerhandbuch

Automatisches gerinnungszeit-messgerät
Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
Control Type
[HR-NM]
Control Lot
46372985102
Temperature
Adjustment

General Quality Control (QC) Functions

The [Control Type] variable function key is used to select the type of control to be
performed. The following control types are available and will be determined by the
test cartridge being used:
The ACTtrac
490-510.
High-Range ACT (HR-ACT) cartridge: Normal (HR-NM) and Abnormal
(HR-AB) controls.
Low-Range ACT (LR-ACT) cartridge: Normal Citrated Whole Blood (CWB)
and Abnormal (LR-AB) controls.
High-Range Heparinase (HR-HTC) cartridge: Normal (HR-NM) and
Heparinase (HTC) controls.
Recalcified ACT (RACT) cartridge: Normal Citrated Whole Blood (CWB) and
Abnormal (RACT-AB) controls.
General Purpose (GPC) Cartridge: Level 1 and Level 2 controls.
The [Control Lot] variable function key is used to enter, remove, or edit the lot
number, expiration date, and range of the selected control. Up to two lots of each
control type may be entered into the database at one time.
The [Temperature Adjustment] variable function key is used to verify the actuator
heat block temperature (see Actuator Heat Block Temperature on page 41).
The [QC Due Status] variable function key displays the status of liquid and
electronic controls, including the date and time that the control tests must be
performed.
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
--->> QC <<---
QUALITY CONTROL MENU
Figure 5. Quality Control Menu Screen
®
electronic control has three settings: 98-102, 190-204, and
Medtronic Confidential
Temp: 37.1°C
QC Due Status
QC Manager Setup
Exit to
Main Menu
English 25
CS010

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis