Herunterladen Inhalt Inhalt Diese Seite drucken

To Change The Lot Number Of The Active Control; To Remove A Control Lot; To Edit The Range Of An Existing Control Lot; Enter A Control Lot Number, Expiration Date, And Range Using The Optional Bar Code Scanner - Medtronic ACT Plus Benutzerhandbuch

Automatisches gerinnungszeit-messgerät
Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
When there are two lot numbers entered for a control type, the following variable
function keys appear: [Toggle Active], [Remove Lot], [Edit Range], and [Exit to
Quality Control Menu].

To change the lot number of the active control:

1. Select [Toggle Active] to move the "
currently being used.

To remove a control lot:

1. Select [Remove Lot].
2. To select the lot to be removed, use the up/down arrows (↑↓) to toggle to the
appropriate lot number.
3. Select [Remove Selected Lot] to delete the selected control lot number.

To edit the range of an existing control lot:

1. Select [Edit Range].
2. To select the lot of the control to be edited, use the arrows to change the
highlighted lot.
3. Enter the range for the control.
4. To confirm the entry, press [Enter].
Enter a Control Lot Number, Expiration Date, and Range
Using the Optional Bar Code Scanner
From the Quality Control Menu:
1. Select [Control Lot].
When there is no or one lot number for a control type:
1. Scan the bar code on the control box. The lot number and expiration date will
automatically populate their respective fields.
2. To enter the range for the control, select [Set Range]. The format for entering
the control range is "lll-hhh," where "lll" is the low end and "hhh" is the high end
of the control range.
3. To confirm the entry, press [Enter].
4. To return to the Quality Control Menu, select [Exit to Quality Control Menu].
When a control type has two lot numbers entered for a control type:
1. To remove one of the lot numbers for the control type, select [Remove Lot].
2. To delete the selected control lot number, toggle to the appropriate lot number
and select [Remove Selected Lot].
3. To enter the new lot number, scan the bar code on the control box. The lot
number and expiration date will automatically populate their respective fields.
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
*
" to the control lot number that is
CS010
English 27

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis