Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
9. Terminate the test by pressing the Stop switch. The actuator heat block
should rotate to the open position.
10. Insert the temperature verification cartridge. After 10 minutes, check both the
temperature indicator on the front panel and the temperature verification
cartridge to confirm that both show a temperature of 37°C ± 0.5°C. Adjust the
temperature, if necessary (see Temperature Verification Cartridge on page 42
for detailed instructions).
11. Insert a floppy disk into the floppy drive.
12. From the Main Menu, select [Transmit Test Results], then [Transmit Unsent
Patient Tests]. The results of the previous test should be sent to the floppy
disk (see Specifications on page 10 for floppy disk requirements, and Data
Entry on page 18 for detailed data entry instructions).
13. If the instrument does not function properly, contact a Medtronic Service
Representative (see Service and Troubleshooting on page 44).

Setup

Medtronic suggests taking the following steps prior to running tests with the ACT
TM
Plus
instrument. Consult your Point of Care Coordinator for details.
1. Set Instrument Parameters (see page 23).
2. Set QC Manager parameters:
a. Enter Instrument Location (see page 30).
b. Enter User IDs (see page 30).
c. Select QC Interval (see page 32).
d. Set QC Lockout to ON (see page 31).
e. Enable Cartridge Type (see page 32).
f. Set Permanent Record (see page 32).
3. Enter Cartridge Lot Information (see pages 21-22).
4. Enter Control Lot Information (see pages 26-27).
5. Run Controls.

Default Operating Parameters

The ACT Plus™ instrument is shipped with the following default parameter
settings:
Patient Parameters
Patient ID
User ID
Cartridge Lot Number
Cartridge Exp Date
Control Lot Number
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
None
None
1111111111
11/11/11
1111111111
CS010
English 13