Herunterladen Inhalt Inhalt Diese Seite drucken

Alerts; Error Messages - Medtronic ACT Plus Benutzerhandbuch

Automatisches gerinnungszeit-messgerät
Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
The passwords do not match.
QC past due.
Test terminated before clot detect.
No User IDs to download.
No valid User IDs in file.
Test requires valid User ID.

Alerts

Bar code scan timeout.
> 10% spread error.
> 12% spread error.
Error - File not found.

Error Messages

Disk error - Disk is faulty.
Serial port error.
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
When changing the password for the QC
Manager, the Confirm Password screen did
not match the Change Password screen.
Reenter the same password in the Confirm
Password screen that was entered in the
Change Password screen.
The QC for this cartridge has passed its
interval. Perform QC tests for this cartridge.
The test stopped before clots were detected
in both channels. Repeat test.
The QC Manager is trying to download User
IDs to the floppy disk, but no User IDs are in
the memory. Add User IDs to the memory
and download to the floppy drive.
The QC Manager is trying to upload User
IDs from the floppy disk, but there are no
User IDs in the file. Download User IDs from
an ACT Plus™ instrument or an ACT Plus™
External Data Manager (EDM) to the floppy
disk, and use that floppy disk to upload them.
The QC Manager has set User Lockout to
ON so that a User ID is required to run a
patient test. Enter a valid User ID in the
Enter ID screen.
A bar code input sequence was detected,
but it did not complete correctly. Rescan the
bar code.
The channels of the controls tested are not
within 10%. Repeat the control test.
The channels of the controls tested are not
within 12%. Repeat the control test.
The ACT Plus™ instrument is trying to read
a file on a floppy disk, but cannot find it.
The ACT Plus™ instrument is trying to read
a floppy disk and cannot read it. Insert a
floppy disk that meets the specification listed
in Specifications on page 10.
An error occurred in accessing the serial
port. Retry transfer.
CS010
English 47

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis