ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
To change the QC Manager Password:
From the User ID Setup screen:
1. Select [QC Mgr Password].
2. Enter up to a 6-digit numeric password.
3. Enter the password a second time to verify.
4. To confirm, press [Enter].
5. Record the new password in a secure location.
Download User IDs to a Floppy Disk
From the User ID Setup screen:
1. Insert a formatted disk into the floppy drive.
2. Select [Download User IDs]. All of the User IDs entered will be downloaded
to the floppy disk.
Upload User IDs from One ACT Plus™ Instrument to
Multiple Instruments
From the User ID Setup screen:
1. Insert a formatted disk containing downloaded User IDs into the floppy drive.
2. Select [Upload User IDs].
User IDs also can be uploaded to the ACT Plus™ instrument using a file generated
from the ACT Plus™ External Data Manager (EDM). Refer to the EDM Instructions
for Use.
Select the Time Interval for Clearing Patient and User IDs
From the QC Manager Setup screen:
1. Select [Clear ID Interval].
2. To view the choices, [None], [5 minutes], [10 minutes], and [20 minutes],
continue to press [Clear ID Interval].
3. To confirm the choice, press [Enter].
Select the QC Lockout Function
From the QC Manager Setup screen:
1. Select [QC Lockout].
2. To view the choices, [ON], [Warning], or [OFF], continue to press [QC
Lockout].
3. To confirm the choice, press [Enter].
ACT Plus™ Operator's Manual
86702-001
Rev 1.0
Medtronic Confidential
CS010
English 31