Herunterladen Inhalt Inhalt Diese Seite drucken

Set The Qc Interval For Liquid And Electronic Controls; To Set The Liquid Qc Interval; To Set The Acttrac® Electronic Qc Interval; Enable Or Disable Test Cartridge Types - Medtronic ACT Plus Benutzerhandbuch

Automatisches gerinnungszeit-messgerät
Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)

Set the QC Interval for Liquid and Electronic Controls

From the QC Manager Setup screen:
1. Select [QC Interval Setup]. The following variable function keys appear:
[Liquid Interval], [ACTtrac

To set the liquid QC interval:

1. Select [Liquid Interval].
2. To view the choices, [None], [8 hours] or [7 days], continue to press [Liquid
Interval].
3. To confirm the choice, press [Enter].
®
To set the ACTtrac
electronic QC interval:
®
1. Select [ACTtrac
2. To view the choices, [None] or [8 hours], continue to press [ACTtrac
Interval].
3. To confirm the choice, press [Enter].

Enable or Disable Test Cartridge Types

From page 2 of the QC Manager Setup screens:
1. Select [Cartridge Enable Setup].
2. To select the desired cartridge, use the arrows on either side of the display of
cartridges to navigate up or down.
3. Select the desired cartridge.
4. To view the choices, [ON] or [OFF], continue to press the desired cartridge.
5. To confirm the choice, press [Enter].

Download All of the Instrument Settings to a Floppy Disk

From page 2 of the QC Manager Setup screens:
1. Insert a formatted disk into the floppy drive.
2. Select [Download Settings].
Set the Permanent Record or the Mode by Which the
ACT Plus™ Instrument Maintains a Record of
Downloaded Data
From page 2 of the QC Manager Setup screens:
1. Select [Perm Record].
2. To view the choices, [None], [Floppy], or [Electronic], continue to press
[Perm Record].