Malfunctions / Troubleshooting
Display symbol(s)
Table 5
Status and error messages
Page 158
Downloaded from
www.Manualslib.com
Error / warning
Electric footrest motor fault
Drive motor fault
Brake fault
Emergency stop
Communication error
(alternating flashing signal)
Battery under voltage
Battery over voltage
manuals search engine
Cause
Faulty cabling or plug connection;
defective actuator
Faulty cabling or plug connection;
defective actuator
Open brake release / defective
brake
Severe fault caused by controller,
manual control device, and / or
drive motor malfunction
Faulty connection between manual
control device and controller /
defective cabling, software or
hardware
Battery deep discharge
Voltage too high (after full charge
and driving downhill)
02/2009
Possible Corrective Action
Check cabling / plug connec-
tion; contact specialist dealer
Check cabling / plug connec-
tion; contact specialist dealer
Close brake release; check
brake (e.g. the cable)
Contact specialist dealer
Check cabling / plug connec-
tion; contact specialist dealer
Charge as soon as possible
Continue driving slowly