Section IX - Specifications

Kendall SCD™ 700 Series Compression System
Safety Standards
Device Classification
Mode of Operation
Ingress Protection
Compression Type
Compression Cycle
Set Pressure
Adjustable Bed Hook
Power Cord Storage
Audible/Visual Errors
Power Cord
Controller Dimensions
Controller Weight
Power Requirements
Battery
Shipping Unit
Shipping Case Dimensions
Shipping Weight
Tubing Set
Operation & Service
Manual
Operating Conditions
Transport & Storage
EN-27
Built to UL60601-1, CSA-C22.2 No. 601.1-M90, CSA C22.2 N0. 60601-1: 2008, JIS
T 0601-2-204, EN60601-1, ANSI/AAMI ES60601-1:2005 and IEC 60601-1-2:2007
Standards UL Classified File # E189131 and E351453
Class I - Equipment Internally Powered, Portable Type BF Applied Parts Not AP
or APG Equipments
Continuous
IP23 (EN60529)
Leg Sleeves: Sequential, Gradient, Circumferential; Foot Cuffs: Uniform
Leg Sleeves: 11 Seconds Compression; Foot Cuffs: 5 Seconds Compression
Decompression time based upon Vascular Refill Detection measurement
Leg Sleeves: 45 mmHg
Foot Cuffs: 130 mmHg
Yes
Yes
Low Pressure, High Pressure, Internal Electronics Malfunction
13 feet long with region specific appropriate cordage and plug
Height: 6.8 inches (17.3 cm)
Width: 7.7 inches (19.6 cm)
Depth: 4.5 inches (11.4 cm) (when placed on a foot board)
Depth: 7.3 inches (18.5 cm) (free standing)
5.0 lbs. (2.3 kg)
100-240 VAC, 50VA, 50/60 Hz
10.8 V, 2200mAh, Lithium Ion pack
Run Time: 6-8 hours
Charge Time: 4 hours (charging only)
Each
11.6 inches (29.4 cm) X 9.25 inches (23.5 cm) X 13.25 inches (33.7cm)
7 lbs. 4 oz. (3.3 kg)
Included, set of two individual assemblies
Included as either CD or Paper Manual
Temperature: 10°C to 40°C
Relative Humidity: 85% Maximum, non-condensing
Atmospheric Pressure: 700 mbar to 1060 mbar
-20°C (-4°F) to 55°C (131°F) If the user suspects that the environment conditions