Herunterladen Inhalt Inhalt Diese Seite drucken

Error Codes - Covidien Kendall SCD 700 Bedienungs Und Servicehandbuch

Sequenzielles kompressionssystem
Vorschau ausblenden Andere Handbücher für Kendall SCD 700:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 74

Error Codes

Error Code
Garment Mismatch Error
A
System High Pressure Error
E1
A
High Pressure (Leg Sleeves)
E2
A
High Pressure (Foot Cuffs)
E2
A
Low Pressure (Leg Sleeves)
Low Pressure (Foot Cuffs)
EN-15
Error Type
User Resettable
B
Manual Reset required
B
User Resettable
B
User Resettable
B
User Resettable
User Resettable
Description
The Garment Detection
procedure has detected
a garment configuration
(Leg or Foot flashing green)
that does not match the
User-selected configuration
(Leg or Foot red).
System pressure has
exceeded 90 mmHg (Leg
sleeve) or 180 mmHg (Foot
Cuff).
Leg Sleeve pressure is
greater than 47 mmHg
for 10 consecutive cycles;
or pressure is above 65
mmHg for 5 consecutive
cycles.
Foot Cuff pressure is
greater than 135 mmHg
for 10 consecutive cycles
or pressure is above 160
mmHg for 5 consecutive
cycles.
Leg Sleeve pressure is
less than 43 mmHg for 10
consecutive cycles.
Foot Cuff pressure is less
than 125 mmHg after 10
consecutive cycles.
Troubleshooting
Press the port
configuration button(s)
to turn the foot selection
on/off depending on
what type of garment(s)
is connected to the
controller. If the proper
garment is selected and
the problem persists have
the controller serviced by a
professional.
Check for kinked tubes or
patient interference with
the garments, like pressing
foot against foot board.
Check for a tight leg
sleeve and adjust fit
appropriately. Also check
for a partially occluded
tube.
Check for a tight foot
cuff and adjust fit
appropriately. Also check
for a partially occluded
tube.
Check for leaks in the
sleeve or the tube
connections.
Check for leaks in the cuff
or the tube connections.
Kendall SCD™ 700 Series

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis