If the patient not detected alert shown in Fig 5 or Fig 6 is displayed, turn the controller off, confirm the leg or foot
garment placement and securement. Then, turn the controller on.
There may be instances in which a patient is present but is not able to be detected by the controller. In this instance,
the white icon in the top corner (Figure 7) will turn blue (Figure 8) but the controller will continue to perform
compression cycles. After two Patient Not Detected Alerts and subsequent power cycles within one hour, the system
will silence the Patient Not Detected Alert. When the icon is blue the Patient Not Detected Alert will be silenced (Off).
The controller will continue to attempt to detect the patient in the background, even though the alert is silenced. If a
patient is subsequently detected, the icon in the top right corner will turn white meaning the controller is now able
to be detect the patient (Patient Detection is Active).
The Patient Detection feature may be disabled in Test Mode (see Test Mode T8). In this instance, there will be no
patient detection icon displayed on the user interface and the unit will not attempt patient detection until the
feature is re-enabled. The therapy time will increment while the device is running and delivering therapy.
EN-9
FIGURE 6 – PATIENT NOT DETECTED ALERT SCREENS (FOOT CUFFS)
-
figure 7
patient detection active
-
figure 8
patient detection inactive
The white icon in upper right hand corner
indicates that patient is actively being detected
by the controller.
On the compliance meter screen, the therapy
time is based on the time that therapy is being
delivered to the patient.
If the sleeve is removed, the controller will
display the Patient Not Detected Alert and
sound an audible alert.
The blue icon in upper right hand corner
indicates that patient is not currently being
detected by the controller. No intervention is
required when the blue icon is displayed, it is
just informational. The controller will continue
to perform compression cycles and attempt to
detect the patient's presence.
On the compliance meter screen, the therapy
time is based on the time that the device is
running.
If the sleeve is removed, the controller will NOT
display the Patient Not Detected Alert and will
NOT sound an audible alert.