Accessing the Compliance Meter Feature

The Compliance Meter can only be accessed when the controller is powered on and is delivering therapy. The
controller will sound a "deny" tone, three quick beeps, at any other time such as immediately after turning the system
on and garment detection is in progress (garments icon blinking).
Note: Use of the Compliance Meter does not halt or otherwise affect the ongoing compression therapy.
• Access the Compliance Meter by pressing the right arrow button. A screen similar to the one in Figure 4 will be
displayed. Pressing it again will return the user to normal operating mode. If the Compliance Meter is accessed,
but no further action is taken, then the system will change the display back to normal operating mode in thirty
seconds.

Reading the Compliance Meter

• The lower left corner of the screen displays numbers and a symbol representing time durations. Typical nurse
shifts are 8, 10, and 12. A full day is 24. The symbol ∑ represents total compliance time since the last reset.
• Pressing the A button (shift selector) allows the user to select a time duration of interest. The selection indicator
moves with each button press.
• Examples:
° To determine the amount of therapy a patient has received over the most recent 8 hours, for example, select
the '8' on the shift selector.
° To determine the amount of therapy a patient received over the most recent 24 hours, for example, select the
'24' on the shift selector.
Note: If the amount of elapsed time has not yet reached the time selected on the shift selector, then the actual
elapsed time will display in the bottom number.
Note: After 30 seconds of inactivity, the Compliance Meter will return to the normal therapy screen.

Resetting the Compliance Meter

1. Press the B button to reset the Compliance Meter. The confirmation screen will display as shown in Fig 9.
2. To confirm the reset, press the A button. A check mark icon displays to confirm your selection.
3. To decline the reset, press the B button. An X icon displays to confirm your selection.
4. After either A or B is pressed, the screen will revert to the Compliance Meter Screen.
5. When the Compliance Meter is reset, the therapy and elapsed time are reset to zero.
6. The compliance Meter continues its operation until it is reset. That is, the therapy time and elapsed time
continue counting until the meter is reset. This may result in inaccurate compliance information for the patient.
However, it is not recommended to reset the meter until it is assigned to a new patient.
Kendall SCD™ 700 Series
Reset icon
Confirm reset
FIGURE 9 - Compliance Reset Screen
Deny reset
EN-10