Applying or releasing the optional kickstand for dialysis scale (Continued)
EN
WARNING (CONTINUED)
Always lower the cot height before you apply the kickstand for increased stability.
Do not apply the kickstand during transport. Keep the kickstand in the retracted position.
Do not use the kickstand as a brake.
Do not apply the kickstand on a sloped surface.
Note: The optional kickstand (6085-102-000) is not compatible with the optional base storage net (6500-160-000).
To apply the kickstand:
1.
Operator 1: Apply the kickstand with your foot.
2.
Operator 2: Lift the foot end of the cot to actuate the kickstand.
3.
Both operators: Make sure that the kickstand is in the forward locked position.
To release the kickstand, Operator 1 lifts the foot end of the cot until both wheels are off the floor. Operator 2 rolls the
cot forward to make sure that the kickstand retracts.

Securing the patient with the G-rated restraint straps

WARNING
Always use all restraint straps to secure the patient on the cot. An unrestrained patient may fall from the cot.
Do not attach restraint straps to the base tubes or cross tubes.
CAUTION
Do not entangle the restraint straps in the base frame when you raise or lower the cot.
Note: Restraint straps are a Type BF applied part.
Secure restraint straps to the cot in the required attachment locations
attachment locations should provide strong anchorage and proper restraint position. Do not allow restraint straps to
1-30
Operation
(Figure 1-10 on page
6086-209-005 REV G
1-31). Restraint strap
www.stryker.com