CPR EMERGENCY RELEASE

When quick access to the patient is needed and the Fowler, Gatch, or Foot end are raised, the bed can be positioned
at 0° by pressing on the CPR pedal . The Fowler, Gatch and Foot end will automatically flatten .
WARNING
The CPR pedal is for emergency use only . When activating the CPR pedal, all people and equipment must be removed
from the area below and around the head, thigh and foot sections of the bed or serious personal injury or equipment
damage could occur .

LOWERING THE FOWLER DURING A POWER FAILURE

If there is a power failure and the backup battery is out, the CPR pedal can be used to partially or completely lower the
Fowler . Simply depress the CPR pedal until the desired angle is reached .

FOLEY BAG HOOKS

The four Foley bag hooks are located on both sides of the bed frame under the foot and seat sections . Two are
centered and isolated .
Note
The Foley bag hooks move when the Fowler is activated . Fowler motion must be locked out when using these hooks to
avoid inadvertent movement of the hooks .

PATIENT RESTRAINT LOCATIONS

The bed has 12 locations on the mattress support for installing patient restraint straps . Ten of them are located on
the mattress support edges directly across from each other and the remaining two are located on the top edge of the
head section .
WARNING
Improperly adjusted restraint straps can cause serious injury to a patient . It is the responsibility of the attending medical
personnel to determine proper use of the restraint straps and restraint strap locations . Stryker is not responsible for the
type and/or use of restraint straps on any of Stryker's products .
CAUTION
Clean Velcro AFTER EACH USE . Saturate Velcro with disinfectant and allow disinfectant to evaporate . (Appropriate
disinfectant for nylon Velcro should be determined by the hospital .)