Installing the defibrillator platform (Continued)
EN
Note: If the defibrillator platform is not attached to the product after you connect both latch hooks, unlatch the tab. To
adjust, loosen or tighten the strap and then push the tab up until it locks into place with an audible click.
Hanging equipment from the equipment hook
Use the equipment hook to hang additional accessories or equipment, such as defibrillators and monitors.
CAUTION
•
Do not load the equipment hook above the safe working load of 35 lb (15.8 kg).
•
Always remove all accessories or equipment from the equipment hook when in the vehicle.
Installing the head extension with pillow
You can install the head extension assembly on the backrest to provide head end support.
To attach the pillow to the head extension, place the support into the flap on the bottom of the pillow. Secure the pillow
to the Velcro® on the bottom of the support.
Notes
•
The head extension with pillow (6100-044-000) is not compatible with the optional equipment hook (6500-147-000)
or optional fowler oxygen bottle holder (6500-241-000).
•
The head extension pillow is a Type BF applied part.
1-40
Operation
Figure 1-30: Equipment hook
6086-209-005 REV G