Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting; Limited Warranty; Liability; Life Of Product - COOK Medical K-MAR-5200 Handbuch

Vorschau ausblenden Andere Handbücher für K-MAR-5200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 57
K-MAR-5200 Instructions for Use

9. Troubleshooting

IMPORTANT NOTE: Should any errors persist, contact your COOK distributor.
1
Error and Alarm Indicator
Vacuum Pump will not turn on.
Displays do not illuminate.
Desired aspiration vacuum is not achieved.
Chime Volume Adjustment does not adjust.
Vacuum Pump displays the Vacuum in wrong
units.
Vacuum stays on when the Foot Pedal is
depressed and released.
Vacuum doesn't stay on when Foot Pedal is
depressed and released.
Vacuum Pump displays an error code in the form
of ER1 to 5.

10. Limited Warranty

COOK AUSTRALIA warrants to the purchasers of this device that at time of manufacture, the product was prepared and tested in accordance with good
manufacturing practices and guidelines specified by the Australian Therapeutic Goods Administration or relevant competent authority.
In the event of product failure under normal use, due to defects in material or workmanship, within a period of one (1) year from the date of purchase, the
product will be repaired, or at Cook's option, replaced, at no charge. This limited warranty does not apply to products subjected to abnormal use or conditions,
improper storage, damaged by accident, misuse or abuse, improper line voltage or to products altered or serviced by anyone other than Cook Australia or its
authorised agent.
The foregoing limited warranty is exclusive and in lieu of all other warranties whether written, oral, expressed or implied. In particular, Cook Australia does not
warrant that the product is suitable for the needs of the purchaser and there are no warranties given as to merchantability or fitness for a particular purpose.
Cook Australia's representations concerning fitness for purpose or suitability for use by any purchaser does not extend beyond those representations set out
in the Cook Australia literature that accompanies the product. Cook Australia assumes that the purchaser is experienced in the use of this device and is able to
judge from his/her own expertise the suitability or otherwise of the product for the intended use. Cook Australia conducts a technical advisory service, which
can be consulted by a purchaser or intended purchaser on an advisory basis.
After one (1) year from the date of purchase, this device will be repaired for a repair charge equal to the cost of parts, labour and transport.
Before returning a product for any reason, please contact your nearest Cook distributor for assistance and instructions.
Cook Australia reserves the right to change or discontinue this product without notice.

10.1 Liability

Because Cook Australia has no control or influence over the conditions under which this device is used, over its method of use or administration, or on handling
of the product after it leaves its possession, Cook Australia takes no responsibility for the results, use and/or performance of the product. Cook Australia expects
that use of the product will be confined to trained and expert users.
In no event will Cook Australia be liable for any direct or indirect damages including incidental, consequential or special damages, arising out of or in connection
with the use or performance of the product.
If the manufacturer provides you with technical documentation, this does not authorise you to perform repairs, adjustments or alterations on the device or
disposables.
No representative of Cook Australia and no vendor or lessor of the product is authorised to change any of the foregoing terms and conditions, and the purchaser
accepts the product subject to all terms and conditions herein, subject always to any contrary provisions which are necessarily implied by stature or law
notwithstanding the within terms and conditions.

10.2 Life of Product

The expected service life is deemed to be seven (7) years. After this time Cook Australia will no longer be responsible for this product.
Source of Error
Power supply cord is not connected. Mains
power switch is not turned on.
The device is in standby mode.
Foot Pedal is defective.
Leak in the Vacuum Line or the Filter is wet.
Power supply cord is not connected. Mains
power is not turned on.
The device is not in standby.
The Display Units setting is set to either
mmHg or kPa.
The Foot Pedal Function is set to Latching.
The Foot Pedal Function is set to Non–
Latching.
Device has an internal fault.
1-17
Elimination of Error
Check the power supply connection. (Refer to
§3.10).
Check that the Standby Indicator is illuminated
orange. Press the Standby Touch-Pad. (Refer to
§3.10).
Disconnect the Foot Pedal from the rear of the
device without applying pressure to the Foot Pedal
and re-connect the Foot Pedal.
Change the Vacuum Line and Filter. (Refer to §7).
Check the power supply connection. (Refer to
§3.10).
The device is functionally active, press the Standby
Touch-Pad to place the device in standby. (Refer
to §3.15).
Set the Display Units setting to desired units, either
mmHg or kPa. (Refer to §3.12).
Set the Foot Pedal Function to Non–Latching.
(Refer to §3.13.3).
Set the Foot Pedal Function to the Latching. (Refer
to §3.13.3).
Contact service agent.
English

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis