K-MAR-5200 Instructions for Use
WARNING: ELECTRIC SHOCK
HAZARD. Internal circuitry is energised
whenever the Vacuum Pump is
connected to mains power irrespective
of whether the device is on or in
standby.
IMPORTANT NOTE: It is
recommended that the Vacuum Pump
be given a pre-operation test before
each operation.
IMPORTANT NOTE: If the
vacuum value is not reached or if the
value begins to decrease again, then
there is a leak. First check the
Disposable Vacuum Line with
Hydrophobic Filter.
IMPORTANT NOTE: If you
should find or suspect deficiencies
in the Vacuum Pump during the
described function control, the device
must not be used until the authorised
service agent has repaired it.
Never use the Vacuum Pump if there
are obvious deficiencies, especially
involving the power plugs or the power
supply connection cables.
Arrange for repair by an authorised
service agent.
4. Installation and Set-up Checklist

3.16 Pre-Operation Test

• Connect the Foot Pedal and the Disposable Vacuum Line with Hydrophobic Filter (K-DVLF-240) to the Vacuum Pump.
• Switch the device on.
• Ensure that the device is set to display in mmHg mode (see § 3.12).
• Use the Vacuum Adjust Touch-Pads to select a vacuum between -120 and -170 mmHg.
• Activate the Foot Pedal. The Vacuum Supplied Indicator lights up. The chime sounds.
• The Vacuum Display may momentarily decrease (e.g from -170 to -160 mmHg). The Pump Motor may then be
heard to accelerate and bring the vacuum level back to the set level, within ±5 mmHg.
• Close the Vacuum Line by kinking, press and hold the Boost Touch-Pad.
• The device should achieve and display a vacuum of -500 mmHg. Note the vacuum is likely to overshoot and may
reach -530 mmHg.
• Release the Boost Touch-Pad and the Vacuum Line.
• Deactivate the Foot Pedal.
• The Vacuum Display should reach the pre-selected value within ±5 mmHg.
The Pre-Operation Test is now successfully completed and the Vacuum Pump is ready for use in the operating room.
Check the following:
All items have been supplied.
The packaging has been safely stored for future use.
All non-sterile items have been removed from plastic covers.
The Power Cord is correct for your region.
The Vacuum Pump has been placed in a suitable location.
The Vacuum Pump has undergone a Pre-Operation Test.
The Disposable Vacuum Line with Hydrophobic Filter have been connected.
The Foot Pedal has been connected.
The Vacuum Pump has been activated.
The Vacuum Display has been set to the desired units.
The vacuum has been adjusted to the desired value.
The Foot Pedal Function has been set to the desired setting.
1-10
English
1