Installation and Commissioning
2.1
Operating environment
Observe the following points to ensure that the device functions properly and has a long service life
The ambient temperature must be10 to 40 °C.
The relative humidity must be below 80 % (without condensation).
The titration head of the sample changer must be able to move freely.
Do not use or store the device in dusty, damp or wet places.
Make sure that no liquids can get into the device and wipe off any liquids on the device immediately.
Protect the device from vibrations, direct sunlight, corrosive gases and strong magnetic and/or electric
fields.
Use only original spare parts and accessories.
2.2
Unpacking and setting up
The device has been individually assembled for you, therefore there may be deviations with regard to the
described scope of delivery and the accessories. Please refer to the enclosed packing list for the exact scope of
delivery. If you have any questions, please contact us directly (for the service address,
operating
manual).
The device and all accessories as well as the peripheral devices have been carefully checked at the factory for
function and dimensional accuracy. Please ensure that even the small additional parts are removed from the
packaging without leaving any residue.
The device can be set up on any flat surface. Do not push any objects under the device.