Herunterladen Diese Seite drucken

Storz TP101 Gebrauchsanweisung Seite 81

Werbung

73
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Table 3 – Test levels for Radiated and Conducted Immunity Tests
Guidelines and manufacturer's declaration – electromagnetic immunity
The TELE PACK + is designed for operation in the electromagnetic environment defined below. The user of the TELE PACK + should make sure that it is
used in such an environment.
Interference immunity
EN/IEC 60601
tests
test level
Conducted RF disturbances
3 V
rms
acc. to IEC 61000­4­6
150 kHz to 80 MHz
Radiated RF disturbances
3 V/m
acc. to IEC 61000­4­3
80 MHz to 2.5 GHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic waves is affected by absorptions and reflections of buildings,
objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed transmitters,
an electromagnetic site survey should be considered. If the measured field strength at the location where the TELE PACK + is used exceeds the above
compliance levels, the TELE PACK + should be monitored to ensure proper function. If abnormal performance is observed, additional measures may be
necessary, such as re­orienting or relocating the TELE PACK +.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Electromagnetic
Compatibility (EMC)
information
Conformity level
Electromagnetic environment – guidelines
Portable and mobile RF communications equipment should be used
no closer to any part of the TELE PACK +, including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
f
3 V
rms
Where P is the rated power of the transmitter in watts [W] according
to the information provided by the transmitter manufacturer and d is
the recommended separation distance in meters [m].
Field strength of fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range
f
3 V/m
h
Interferences may occur in the vicinity of equipment marked with the
following symbol:
7
Indicaciones sobre
compatibilidad
electromagnética (CEM)
, should be less than the compliance
a
b
.
80 MHz to 800 MHz
800 MHz to 2.5 GHz

Werbung

loading