CAUTION
The mirror found in the tip is a delicate optical component that should be handled with care to ensure optimal scan quality. Be careful not to
scratch or smudge it as any damage or blemishes may affect the data acquired.
Make sure to always wrap the tip before autoclaving. If you autoclave an exposed tip, it will cause stains on the mirror, which cannot be removed.
Check the autoclave manual for more information.
Tips that have been cleaned, disinfected, and sterilized must remain sterile until they are used on the patient.
Medit is not responsible for any damage, such as distortion of the tip, that occurs during cleaning, disinfection, or sterilization operations that are
not following the guidelines above.
3.2.3
Mirror
The presence of impurities or smudges on the tip mirror may lead to poor scan quality and an overall poor scanning experience. In such situations, clean
the mirror following the steps below :
Disconnect the scanner tip from the i700 wireless handpiece.
Pour alcohol on a clean cloth or cotton-tipped swab and wipe the mirror. Make sure to use alcohol that is free of impurities or it may stain the mirror.
You can use either ethanol or propanol (ethyl-/propyl alcohol).
Wipe the mirror dry using a dry, lint-free cloth.
Make sure the mirror is free of dust and fibers. Repeat the cleaning process as necessary.
3.2.4

Handpiece

After treatment, clean and disinfect all other surfaces of the i700 wireless handpiece except for the scanner front (optical window) and end (air vent hole).
Cleaning and disinfecting must be done with the device turned off. Use the device only after it is completely dry.
The recommended cleaning and disinfecting solution is denatured alcohol (ethyl alcohol or ethanol) – typically 60 – 70% Alc/Vol.
The general cleaning and disinfecting procedures are as follows :
Turn off the device using the power button.
Unplug all the cables from the wireless hub.
Clean the filter on the front end of the i700 wireless handpiece.
If alcohol is poured directly into the filter, it may seep inside the i700 wireless handpiece and cause a malfunction.
Do not clean the filter by pouring alcohol or cleaning solution directly into the filter. The filter must be gently wiped with a c otton or soft
cloth moistened with alcohol. Do not wipe by hand or apply excessive force.
Medit is not responsible for any damage or malfunction that occurs during cleaning that does not follow the guidelines above.
Put the cover on the front of the i700 wireless handpiece after cleaning the filter.
Pour the disinfectant onto a soft, lint-free, and non-abrasive cloth.
Wipe the scanner surface with the cloth.
Dry the surface with a clean, dry, lint-free, and non-abrasive cloth.
CAUTION
Do not clean the i700 wireless handpiece when the device is turned on as the fluid may enter the scanner and cause malfunction.
Use the device after it is completely dry.
Chemical cracks may appear if improper cleaning and disinfecting solutions are used during cleaning.
3.2.5

Other Components

Pour the cleaning and disinfecting solution onto a soft, lint-free, and non-abrasive cloth.
Wipe the component surface with the cloth.
Dry the surface with a clean, dry, lint-free, and non-abrasive cloth.
CAUTION
Chemical cracks may appear if improper cleaning and disinfecting solutions are used during cleaning.
3.3

Disposal

CAUTION
The scanner tip must be sterilized before disposal. Sterilize the tip as described in the section "3.2.1 Reusable Tip – Sterilization."
Dispose of the scanner tip as you would any other clinical waste.
Other components are designed to conform with the following directives:
RoHS, Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment. (2011/65/EU)
WEEE, Waste Electrical and Electronic Equipment Directive. (2012/19/EU)
3.4

Battery Storage

Put it in a pack or a box and store indoors in a cool environment out of direct sunlight.
Store the battery in a dry place with an ambient temperature of from -20°C to +30°C (-4°F to +86°F).
If not used for a long time, the battery self-discharge may accelerate and slip into sleep mode. To minimize the effect of deactivation,
store the battery pack in between+10°C - +30°C (+50°F to +86°F).
When charging for the first time after long-term storage, the battery capacity may decrease due to pack deactivation. Restore the battery through
several full charge/discharge cycles.
If the battery is stored for more than 6 months, it should be charged at least once every 6 months to prevent the decrease in shelf life due to self-
discharge.
CAUTIONS: Battery Safety Specifications
Safety Specifications
Overcharge
Over-Discharge
Sense Voltage
Steady Voltage
Sense Delay
Sense Voltage
Steady Voltage
Sense Delay
4.225 V ± 0.020
4.025 V ± 0.03
1.0 s ± 0.2
2.50 V ± 0.035
2.90 V ± 0.50
64 ms ± 12.8
Intraoral Scanner i700 wireless
11