The installation will not be processed while the i700 wireless system is connected to a PC. Please be sure to disconnect the i700 wireless USB 3.0
cable from the PC before the installation.
2.2.3
Medit Scan for Clinics User Guide
Please refer to the User Guide of Medit Scan for Clinics: Medit Scan for Clinics > Menu > User Guide.
CAUTION
Equipment maintenance should only be carried out by a Medit employee or Medit-Certified company or personnel.
In general, users are not required to perform maintenance work on the i700 wireless system besides calibration, cleaning, and sterilization.
Preventive inspections and other regular maintenance are not required.

3. Maintenance

3.1

Calibration

Periodic calibration is required to produce precise 3D models. You should perform calibration when :
T he quality of the 3D model is not reliable or accurate when compared to previous results.
Environmental conditions such as temperature have changed.
The calibration period has expired.
You can set the calibration period in the Menu > Settings > Calibration Period (Days).
T he calibration panel is a delicate component.
Do not touch the panel directly. Check the calibration panel if the calibration process is not performed properly. If the calibration panel is
contaminated, please contact your service provider.
T he calibration panel is a delicate component.
You can set the calibration period via Menu > Settings > Calibration Period (Days). The default calibration period is 14 days.
3.1.1

How to Calibrate i700 wireless

Turn on the i700 wireless and launch the Medit Scan for Clinics.
Run the Calibration Wizard from Menu > Settings > Calibration.
Prepare the calibration tool and the i700 wireless handpiece.
Turn the dial of the calibration tool to position
Put the i700 wireless handpiece into the calibration tool.
Click "Next" to start the calibration process.
When the calibration tool is mounted properly in the correct position
When data acquisition is complete at position
Repeat the steps for positions
When data acquisition is complete at the
3.1.2

Auto Calibration Tool (Sold Separately)

The i700 wireless auto calibration tool accessory can be purchased separately. This handy auto calibration tool will automatically perform calibration by
calibrating the i700 wireless handpiece without having to turn the calibration dial. Please refer to Medit Scan for Clinics for details.
3.2

Cleaning, Disinfection, Sterilization Procedure

3.2.1
Reusable Tip – Sterilization
The reusable tip is the part that is inserted into the patient's mouth during scanning. The tip is reusable for a limited number of times. The tip needs to
be cleaned and sterilized between patients to avoid cross-contamination.
The tip should be cleaned manually using a disinfecting solution. After cleaning and disinfecting, inspect the mirror inside the tip to
ensure there aren't any stains or smudges.
Repeat the cleaning and disinfection process if necessary. Carefully dry the mirror using a paper towel.
Insert the tip into a paper sterilization pouch and seal it, making sure that it is airtight. Use either a self-adhesive or heat-sealed pouch.
Sterilize the wrapped tip in an autoclave with the following conditions:
Sterilize for 30 minutes at 121°C (249.8°F) at gravity type and dry for 15 minutes.
Sterilize for 10 minutes at 135°C (275°F) at gravity type and dry for 30 minutes.
Sterilize for 4 minutes at 134°C (273.2°F) at pre-vacuum type and dry for 20 minutes.
U se an autoclave program that dries the wrapped tip before opening the autoclave.
Scanner tips can be re-sterilized up to 150 times and thereafter must be disposed of as described in the disposal section.
Autoclave times and temperatures may vary depending on the autoclave type and manufacturer. For this reason, it may not be able to meet the
maximum number of times. Please refer to the user's manual of the autoclave manufacturer you are using to determine whether the required
conditions are met.
3.2.2
Reusable Tip – Cleaning and Disinfection
Clean the tip immediately after use with soap water and a brush. We recommend using a mild dishwashing liquid. Make sure the mirror of the
tip is completely clean and stain-free after cleaning. If the mirror appears stained or foggy, repeat the cleaning process and rinse thoroughly with
water. Dry the mirror carefully with a paper towel.
Clean with Caviwipes according to the following conditions. Please refer to the Caviwipes instruction manual for proper use.
Caviwipes : disinfect for 3 minutes and dry for 5 minutes.
Caviwipes-1 : disinfect for 1 minute and dry for 5 minutes.
Caviwipes-2 : disinfect for 2 minutes and dry for 5 minutes.
Disinfect the tip using Wavicide-01 for 45 to 60 minutes. Rinse the tip thoroughly. Please refer to the Wavicide-01 Solution instruction
manual for proper use.
Remove the tip from the used solution and rinse thoroughly after cleaning and sterilization.
Use a sterilized and non-abrasive cloth to dry the mirror and the tip gently.
10
User Manual
.
1
, turn the dial to the next position.
1
and the
8
2
LAST
position, the system will automatically calculate and show the calibration results.
LAST
, the system will automatically acquire the data.
1
position.