3. Select the measuring mode
Access the EXPRT-CONFIG-MeasMd menu using the arrow and Set pushbuttons
HiSpd: max. length = 2,005 mm, response time < 400 ms
HiAcc: max. length = 6,005 mm, response time < 2,800 ms, more stable measured values,
recommended for liquids with low DK values and where TrsHld is < 70
4. Static sources of interference in the tank
Static sources of interference in the tank generated by tubes, beams, couplings, or a cleaning
ball are taught into the system as standard.
Access the EXPRT-CONFIG-CalRng menu using the arrow and Set pushbuttons.
The following information applies:
Teach-in depth starts from the LFP process connection
Teach-in depth should cover all interference signals
Maximum teach-in depth (recommended) = probe length
Set the value range between 95 and 6,005 mm
If the tank cannot be emptied completely, the CalRng teach-in depth must be adapted accord-
ingly.
The level must be at least 200 mm below the end of the probe
5. Run the AutCal function
Access the AutCal menu using the arrow and Set pushbuttons.
The following information applies: The probe must not be covered with liquid in the CalRng
set in step 4 (teach-in depth + 200 mm).
Use the Set pushbutton to confirm the AutCal menu item and then use it again to confirm the
security question Ok?.
The AutCal function is confirmed with !CalOK.
6. Analyze the signal quality
The signal quality can be analyzed when the device is installed ("5.6 Evaluating signal quality")
In the event of problems:
Reduce the value in the EXPRT-CONFIG-TrsHld menu
Set the parameter in the EXPRT-CONFIG-MeasMd menu to HiAcc
Switch on the filter in the Set filters menu
Reduce the parameter in the EXPRT-CONFIG-MaxCol menu
7. Configure the filter (see Chapter "5.2 Filtering measured values")
8. Maximum change of level/plausibility check (see Chapter "5.2 Filtering measured values")
9. Configure the outputs (see Chapter 3 "Configuring the switching outputs" and "4 Configuring the
analog output")
Notes
Please use the foam commissioning instructions for applications with foam.
The sensor automatically quits expert mode after 5 minutes of inactivity on the display.
Configuration (AutCal) does not take place in the following processes:
Changing the probe length
Changing the measuring mode
Changing the teach-in depth
If you are encountering problems, see Chapter 7 "Troubleshooting".