Activate the QA analog output
Set the parameter in the QAMENU-SimCur or SimVol menu to the desired signal value.
SimCur for current output
SimVol for voltage output
Note: The simulation is automatically deactivated if the supply voltage is interrupted.
Simulating the level
Even if there is no liquid in the container yet, it is possible to select a filling level in the menu in
order to test the sensor configuration. When simulating a level value, all outputs on the LFP are set
according to the defined configuration. The function should not be selected until a configuration is
complete.
Configuration
Set parameter to the desired filling level as a % in the SimLev menu
Note:
The level simulation refers to the probe length or container level (probe length + offset) if an
offset is configured (see Chapter 5.10 "Setting the offset")
The simulation is only active when there are no error messages. The simulation is automatically
deactivated if the supply voltage is interrupted.
Parameter selection
SimOff: Off
Fill level 0%
Fill level 25%
Fill level 50%
Fill level 75%
Fill level 100%

5.4 Configuring the probe length

Log into expert mode (see Chapter 5.1 "Expert mode")
Access the EXPRT-CONFIG-Length menu using the arrow and Set pushbuttons.
Enter the probe length in the Length menu. Please note the defined probe length in Chapter 9.8
"Dimensional drawings"
Note:
HiSpd: max. length = 2,005 mm, response time < 400 ms
HiAcc: max. length = 6,005 mm, response time < 2,800 ms

5.5 Teaching in static interference signals

Static interference signals in the tank generated by tubes, beams, couplings, or a cleaning ball can
be taught-in. The probe length provides the value for the teach-in depth.
Log into expert mode (see Chapter 5.1 "Expert mode")
Access the EXPRT-Config-CalRng menu using the arrow and Set pushbuttons.
Set the value range between 95 and 6,005 mm
86
8014825/YE12/2014-09-15 © SICK AG • Level sensors • Germany • Subject to change without notice