LFP Cubic
Error
Cause
!Range
The maximum allowable measuring
range was exceeded. Measurement in
this configuration is not possible.
!Cable
The coaxial cable is damaged/faulty
The coaxial cable length was config-
ured incorrectly
Display only
DspVal menu parameter
shows RUN.
is set to QaBarG and the level
Otherwise the
is below QALOW
display is empty.
Display off
Temperature too high
Temperature is too low
No supply voltage
!Err[xx]
System error
!ErM[xx]
!ErI[xx]
!ErO[xx]
NVFail
Memory error
7.2 Operating the display
Error
Cause
The menu item
QxMENU/OUx is not
SPx/RPx is not
configured to Qx-Hno or Qx-Hnc
displayed
The menu item
QxMENU/OUx is not
FHx/FLx is not
configured to Qx-Fno or Qx-Fnc
displayed
QAFAIL is not
The analog output Qa is in voltage
displayed
mode (QATYP = 0 to 10 V)
SimVol is not
The analog output Qa is in current
displayed
mode (QATYP = 4 to 20 mA)
SimCur is not
The analog output Qa is in voltage
displayed
mode (QATYP = 0 to 10 V)
EXPRT-Config-...
Correct password not entered
is not displayed
8014825/YE12/2014-09-15 © SICK AG • Level sensors • Germany • Subject to change without notice
7 Troubleshooting
Solution
Reduce probe length and/or coaxial cable
length (see Chapter "5.7 Changing the coaxial
cable length")
Replace the coaxial cable
See Chapter 5.7 "Changing the coaxial cable
length"
Change QALOW or DspVal
Reduce the temperature
Increase the temperature
Connect sensor correctly
The device is faulty and must be replaced.
Replace the sensor.
The device is faulty and must be replaced.
Replace the sensor.
Solution
Perform configuration of Qx
(see Chapter 3 "Configuring the switching
outputs")
Perform configuration of Qx
(see Chapter 3 "Configuring the switching
outputs")
Perform configuration of Qa
(see Chapter 4 "Configuring the analog
output")
Perform configuration of Qa
(see Chapter 4 "Configuring the analog
output")
Perform configuration of Qa
(see Chapter 4 "Configuring the analog
output")
See Chapter 5.1 "Expert mode"