Fault detected: The Motor will not turn off
Possible cause
Keyboard and/or
footswitch damaged
Broken connections inside.
PCB damaged
Wrong voltage.
Fault detected: The pump is not working properly
Possible cause
Rotor is blocked.
Irrigation cassette not
correctly inserted.
Tube damaged.
Pump damaged/blocked.
PCB damaged.
XII – ELECTROMAGNETIC COMPATIBILITY
Important: The power cord, micromotor cord, and footswitch cord must be kept apart.
The I-
S
™ requires special precautions to be taken with regard to electromagnetic compatibility. It must be
URGE
installed and prepared for use as described in chapter IV.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere with the I-
S
™. The recommended separation distances in this paragraph must therefore be complied with.
URGE
The I-
S
™ must not be used near to or on top of another device. If this cannot be avoided, its operation under
URGE
the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC
consequence of increasing the emissions or decreasing the immunity of the I-
To avoid
possible
of similar nature near I- S
radiation law.
This device has been tested and found to comply with the emissions requirements of IEC 60601-1-2.
These requirements provide reasonable protection against harmful electromagnetic interference in a
typical medical installation. However, high levels of radio-frequency (RF) emissions from electrical
devices, such as cellular phones, may disrupt the performance of this device. To mitigate disruptive
electromagnetic interference, position this device away from RF transmitters and other sources of
electromagnetic energy.
Risks from
influenced by electromagnetic fields. Patients and users must inquire about implanted systems before
using the motors and check the use. Do not bring the motors close to the systems. Take suitable
emergency precautions and immediately react to any changes in health.
Solution
Try to turn off the motor by pressing the [Motor ON/OFF] button on the keyboard or by using
the foot pedal.
If the motor will not turn off, immediately discontinue use of the device and contact
®
SATELEC
after-sales service.
Immediately discontinue use of the device and contact
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another electrical outlet.
Solution
Disconnect power to the central unit. Check to see if the rotor of the peristaltic pump rotates
correctly and is not blocked by any internal or external objects. Remove any visible blockage if
possible. Otherwise, contact
To verify the proper tube connection follow the instructions on Chapter VIII – USING THE
DEVICE.
Check for damage along the tube, kinks, partial occlusions, etc. Arrange for immediate
replacement of the tube.
Immediately discontinue use of the device and contact
risks of electromagnetic interferences do not use other electro-medical instruments
™. The unit complies with the current normative electromagnetic
URGE
electromagnetic
fields: The functions of implanted systems (such as pacemakers) can be
®
SATELEC
after-sales service.
®
as replacement parts may have the
®
SATELEC
after-sales service.
®
SATELEC
after-sales service.
S
™.
URGE
21