page 13): This reduces the nominal pressure by 5 mm Hg within the whole
selectable range.
• Keep the "Pressure (&)" key depressed (see (18) Fig. 5-1 Front of the device,
page 13): This reduces the nominal pressure in increments of 10 mm Hg within
the whole selectable range.
• Briefly press the "Pressure (%)" key (see (16) Fig. 5-1 Front of the device,
page 13): This raises the nominal pressure by 5 mm Hg within the whole
selectable range.
• Keep the "Pressure (%)" key depressed (see (16) Fig. 5-1 Front of the device,
page 13): This raises the nominal pressure in increments of 10 mm Hg within
the whole selectable range.
6.5

Presetting Suction with Opened Shaver (Standard Mode)

The max. opened shaver suction setting is 3 l/min.
The min. opened shaver suction setting is 0.25 l/min.
Factory setting = 0.25 l/min
Briefly press the keys to change the value in increments of 0.25.
Keeping the keys depressed longer than 1.5 seconds scrolls the displayed values
in increments of 0.25.
6.6

Presetting Max. Flow ("Inflow Only Mode")

The suction keys act as "flow limiters" while in "Inflow Only Mode" with the shaver
opened. The displayed value indicates the max. inflow.
The maximum setting is 2.5 l/min.
• Briefly press the "Inflow Limit (&)" key (see (22) Fig. 5-1 Front of the device,
page 13): This reduces the inflow by 0.25 l/min across the entire selectable
range.
• Keep the "Inflow Limit (&)" key depressed (see (22) Fig. 5-1 Front of the device,
page 13): This reduces the inflow in 0.25 l/min increments across the entire
selectable range.
• Briefly press the "Inflow Limit (%)" key (see (20) Fig. 5-1 Front of the device,
page 13): This increases the inflow by 0.25 l/min across the entire selectable range.
• Keep the "Inflow Limit (%)" key depressed (see (20) Fig. 5-1 Front of the device,
page 13): This raises the inflow in increments of 0.25 l/min within the whole
selectable range.
Preparing the Device for Surgery
NOTE
US
21