Turn plug of the wired remote control so that red dot points up.
(2)
1. Connect wired remote control to connection socket at front of device (See Fig.
9-2 page 33).
2. All device functions can be controlled with the remote control.
• Flow
Increasing/reducing shaver suction (1)
• Pressure
Increasing/reducing nominal pressure (2)
• Wash
Rapid change of the flow ('Wash' function) (3)
• Start/Stop
Starting/stopping irrigation process (4)
Please follow the instructions in chapter 12.4 Cleaning, Disinfecting, and
Sterilizing Autoclaveable Wired Remote Control, page 42 for cleaning and
autoclaving.
Fig. 9-2
(1)
(4)
(3)
Device Options
CAUTION
Wired remote control
US
33