Error and Warning Messages

The status field of the display depicts the error and warning messages. In the case of an error message, the display shows
alternating: Device defect / Error message
Error message
Cause
Call service
The
device
operational.
Electronics
Electronic error
defective
detected
Sensor defective
Pressure sensor
defective
Motor defective
Fault of the motor
detected
Keyboard
The keyboard is
defective
defective
Calibration Error
The device is not
calibrated
Warning
Cause
Messages
Critical
The actual pressure
Overpressure
exceeds approx.
250 mm Hg for longer
than 5 s.
Troubleshooting
is
not
Turn the device off and back on using the ON/OFF switch after approx.
10 seconds have expired.
Should the error message appear again, the device is defective. Make sure the
device can no longer be operated until a qualified service technician conducts
the appropriate tests and repairs.
Turn the device off and back on using the ON/OFF switch after approx.
10 seconds have expired.
Should the error message appear again, the device is defective. Make sure the
device can no longer be operated until a qualified service technician conducts
the appropriate tests and repairs.
Turn the device off and back on using the ON/OFF switch after approx.
10 seconds have expired.
Should the error message appear again, the device is defective. Make sure the
device can no longer be operated until a qualified service technician conducts
the appropriate tests and repairs.
Turn the device off and back on using the ON/OFF switch after approx.
10 seconds have expired.
Should the error message appear again, the device is defective. Make sure the
device can no longer be operated until a qualified service technician conducts
the appropriate tests and repairs.
Turn the device off and back on using the ON/OFF switch after approx.
10 seconds have expired.
Should the error message appear again, the device is defective. Make sure the
device can no longer be operated until a qualified service technician conducts
the appropriate tests and repairs.
Call an authorized service technician to calibrate the device. The device
cannot be used and has to remain inaccessible for use until calibrated.
Troubleshooting
Reduce the intra-articular pressure by opening the drainage stopcock on the
instrument.
The device starts automatically after reducing the pressure below nominal
pressure.
Error and Warning Messages