Initial Use of the Device
4.2.7 Recommended Safety Distances between Portable and Mobile HF Telecommunications Devices and the
US
FLOCONTROL ARTHROSCOPY PUMP
Recommended safety distances between portable/mobile HF telecommunications devices and the FLOCONTROL
ARTHROSCOPY PUMP
The FLOCONTROL ARTHROSCOPY PUMP is intended for use in an electromagnetic environment where RF interference
quantities are controlled. The user/operator of the FLOCONTROL ARTHROSCOPY PUMP can contribute to lowering
electromagnetic emissions by complying with the minimum distance between portable and mobile RF telecommunications
devices (transmitters) and the FLOCONTROL ARTHROSCOPY PUMP - depending on the output power of the communication
device listed below.
Rated output of the
transmitter [W]
0.01
0.1
1
10
100
The safety distance d in meters [m] for transmitters with a max. rated output not listed in the table above can be calculated
by applying the corresponding formula in the respective column. P is the max. rated output of the transmitter in watts [W] as
specified by the transmitter manufacturer.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These guidelines may not be applicable in all cases. The distribution and spread of electromagnetic quantities differs
depending on the absorption and reflection of buildings, objects, and people.
12
Safety distance based on the transmitting frequency [m]
150 kHz to 80 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 1.2 P
0.23
0.73
2.3
7.3
23