Special care is necessary when servicing, maintaining, and storing the device and
its accessories to maintain the functionality of the device and its accessories.
The reprocessing instructions are provided in accordance with AAMI ST79,
AAMI ST81, AAMI TIR12, ISO 17664, and ISO 17665. While they have been
validated by the manufacturer as being capable of preparing the devices
intended for re-use properly, it remains the responsibility of the processor to
ensure that the reprocessing, as actually performed (using equipment, materials,
and personnel in the reprocessing facility), achieves the desired result. This
normally requires validation and routine monitoring of the process. The
manufacturer recommends users observe these standards when reprocessing
medical devices.
•To minimize the risk of infection, clean and sterilize all components intended
for re-use prior to the first use and every use thereafter.
•Ensure all components are prepared as described in this chapter prior to
initiating the reprocessing.
•Drying time depends on several variables, including: altitude, humidity, type of
wrap, preconditioning, size of chamber, mass of load, and placement in
chamber. Users must verify that drying time set in their autoclave yields dry
surgical equipment.

12.1 Mechanical Component Check

Make sure the device is set up properly and securely.

12.2 Cleaning the Device

1. Turn off the device using the ON/OFF switch.
2. Remove the power cable.
3. Use a surface disinfectant with a non-alcohol base. The concentration of the
Do not sterilize the device.
NOTE

12.3 Cleaning, Disinfecting, and Sterilizing Secretion Container

Clean, disinfect, and sterilize the secretion container according to the
instructions provided by the manufacturer of the secretion container. Check the
secretion container for cracks, dull, or brittle areas before use.
12.4 Cleaning, Disinfecting, and Sterilizing Autoclaveable Wired
Operate the wired remote control only if dry, closed and sterilized.
12.4.1 Cleaning the Autoclaveable Wired Remote Control
Recommended cleaners:
Slightly alkaline or enzymatic cleaners.
Use a brush.
Follow the instructions of the manufacturer concerning concentration and
Care and Maintenance
disinfectant depends on the information provided by the manufacturer of
the disinfectant. Wipe the surface of the device with a soft cloth moistened
with the disinfectant.
Make sure moisture does not enter the device.
Remote Control