ecommenDeD seParation Distances between Portable anD mobile
245
the moDel
The model 245 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled� The customer or the user of the model 245 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the model 245 as recommended below,
according to the maximum output power of the communications equipment�
r
ateD maximum
outPut Power oF
transmitter
w
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer�
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies�
NOTE 2 These guidelines may not apply in all situations� Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people�

PHYSICAL CHARACTERISTICS

WARMING UNIT
4�5 in� (11 cm) high x 7�5 in� (19 cm) wide x 10 in� (25 cm) long; wt�: 7 lb� 7 oz� (3�4 kg)
CERTIFICATIONS
IEC/EN 60601‑1; IEC/EN 60601‑1‑2; CAN/CSA‑C22�2, No�601�1
CLASSIFICATION
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601‑1; CAN/ CSA‑C22.2,
No.601.1; ANSI/AAMI ES60601‑1:2005 CSA‑C22.2 No. 60601‑1:08; Control No.4HZ8
Classified under IEC 60601‑1 Guidelines (and other national versions of the guidelines) as Class
I, Type BF, Ordinary equipment, Continuous operation� Not suitable for use in the presence of
flammable anesthetic mixtures with air or with oxygen or nitrous oxide� Classified by Underwriters
Laboratories Inc� with respect to electrical shock, fire and mechanical hazards only, in accordance
with IEC/EN 60601‑1 and Canadian/CSA C22�2, No� 601�1 Classified under the Medical Device
Directive as a Class IIb device�
Ranger™ Blood/Fluid Warming System Model 245
s
eParation Distance accorDing to Frequency oF transmitter
150
h
80 mh
k
z to
80 mh
800 mh
z
d = 1.2 √ P
d = 1.2 √ P
0�12
0�12
0�37
0�37
1�17
1�17
3�69
3�69
11�67
11�67
rF
communications equiPment anD
m
800 mh
z to
z to
d = 2.3 √ P
z
0�23
0�74
2�30
7�37
23�30
2,5 gh
z