4.4.2 Speed and Angle Menu

Speed and Angle menu:
The "rocking process" can be started and stopped in both menus. The controller
regulates the rocking rate as well as the angle for the desired target values.
Press "Set" to access the keypad page and enter target values.
Speed control range: 8–42 rocks
Angle control range: 4.0–10.0 °deg
1 rock corresponds to one movement from the center to the front position – then to
the back position – and then back to the center.
The "Sample Position" function automatically moves to the sampling position.
When this button is pressed, the platform moves to a suitable 10° forward position.
At the same time, the heating is turned off to avoid inhomogeneous temperature
distribution. When the button is pressed again, the process continues. The Sample
function is also equipped with an automatic safety timer. Once the time set in the
Sampling Time menu
"4.4.7.5 Sampling Time" on page 38 expires, the rocker turns
t
back on automatically.