Herunterladen Inhalt Inhalt Diese Seite drucken

Temperature Range; Serial Interface; Housing Material; Dimensions | Weight - Sartorius Stedim Biotech BIOSTAT RM 20 basic Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1.4.9 Temperature Range

1.4.10 Serial Interface

1.4.11 Housing Material

1.4.12 Dimensions | Weight

1.4.13 Power supply |
Power Disconnect
1.4.14 Room Temperature
Range

1.5 Bag Holder

1.5.1 Properties, Specifications

Temperature input|Converter + 15 ... + 40 °C, ± 0.2 °C
Minimal temperature depends on the ambient temperature
1 + RS-232
Optional: Profibus DP, Ethernet
Frame: Stainless steel
Outside housing: ABS (UL compliant)
Incl. Bag Holder 20 and hood for Bag Holder 20:
W + D + H 765 + 600 + 400 mm | 30.0 kg
Incl. Bag Holder 50 and hood for Bag Holder 50:
W + D + H 1085 + 600 + 450 mm | 31.3 kg
Power switch on the back of the device. Automatic voltage selection.
100 – 240 VAC | Fuse 6.3 AT
Caution!
Make sure that the power supply is equipped with a ground lead (IEC 61001-1).
Never use the device without a ground lead
0°C to + 40 °C
The bag holder is a platform that holds the Cultibag during operation.
The bag holder includes both heating elements and the bag is attached to both sides
using the lever. Bags can be replaced at any time without the need for tools. The bag
holder is made from ABS and comes in a 20 and 50 version.
Bag Holder Dimensions
in mm |
Weight
Holder 20 765 + 600 + 60 |
5.5 kg
Holder 50 1085 + 600 + 60 |
7.8 kg
* l = left side, r = right side, c = center
** "Optical" and "Perfusion" CultiBags RM require a higher min. volume due to the
integrated single-use sensors.
The BIOSTAT
RM 20 | 50 basic comes standard with a dual temperature
®
measurement and control system. If the rocker is used with a BIOSTAT
Tower "Single" version, only the left temperature measurement function can be
used. A Twin (single rocker) version tower is required for dual bag control.
CultiBag RM
Number of
basic, optical or
mountable
perfusion
bags
CultiBag RM 2L
1 (l*) or 2 (r*)
CultiBag RM 10L
1 (l*) or 2 (r*)
CultiBag RM 20L
1 (c*)
CultiBag RM 50L 1 (c*)
Min.
Max.
operating
operating
volume
volume
in the
in the
CultiBag
CultiBag
RM
RM
[Liter]**
[Liter]**
0.15
1
0.5
5
1
10
5
25
RM DCU
®
Design and Function
15

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Biostat rm 50 basic

Inhaltsverzeichnis