3.3.3 Power Supply Connection

3.3.4 Connecting the Com-
pressed Air and CO
Supply
2

3.3.5 Device Startup

22
Initial Startup
1. Connect the power cable to the rocker and to the power source as shown in
the figure. Please also observe the note in section
page 19.
1. If you are using a compressed air supply, connect it to the "Process Air Inlet"
connection on the back of the basic unit.
2. Connect the CO
source to the corresponding input of the rocker.
2
3. Turn on the pressure for the gasses being used according to the info in section
"1.4.3 Air and CO2 Supply (only with optional aeration module)""1.4.3 Air and
t
CO
Supply (only with optional aeration module)" on page 14.
2
Please note that the device begins operating again automatically using the
set parameters after a power outage.
This function can be activated/deactivated in the System menu under
"Automatic System Restart."
y Make sure that the device has been set up in accordance with the steps listed
above and that all components have been properly installed.
y The device is now operational.
y The "0-1" switch on the back left of the device must be used to turn the device on/
off and/or disconnected it from the power supply.
Warning to avoid hand injuries!
Note that inserting your hands between the bag holder and the rocker is
prohibited during operation. Failure to comply with this may result in material
damage and personal injury.