4.4.7.3 Setting the Bag Size

4.4.7.4 Selecting a
Control Mode
36
Menu Operation
The bag configuration is defined here and displayed on each screen.
The following configurations are available:
1 + 1L:
Only Heater1 (left heating panel) in operation,
Pt100_1 and Temp_1 controller in operation
1 + 2L:
Only Heater1 (left heating panel) in operation,
Pt100_1 and Temp_1 controller in operation
Only Heater1 (left heating panel) in operation,
1 + 10L:
Pt100_1 and Temp_1 controller in operation
1 + 20L:
Heater1 and 2 (left and right heating panel) in operation,
Pt100_1 and Temp_1 controller in operation
1 + 50L:
Heater1 and 2 (left and right heating panel) in operation,
Pt100_1 and Temp_1 controller in operation
Pt100_1 and Temp_1 controller controls Heater1 (left heating panel),
2 + 1L:
Pt100_2 and Temp_2 controller controls Heater2 (right heating panel)
2 + 2L:
Pt100_1 and Temp_1 controller controls Heater1 (left heating panel),
Pt100_2 and Temp_2 controller controls Heater2 (right heating panel)
2 + 10L:
Pt100_1 and Temp_1 controller controls Heater1 (left heating panel),
Pt100_2 and Temp_2 controller controls Heater2 (right heating panel)
The Control mode via the CPU can be selected here.
In this case, the display is selected as the master, i.e. all commands and parameters are
entered via the display as well as all alarms and warning messages.
When you switch the "ON" button, the DCU Tower is activated as the master.
After 3 seconds, it automatically switches to the next page.