1.7 Hood for Bag Holder 20 or 50

1.7.1 Dimensions

1.8 Surface Temperature
Sensor
The hood is a safety cover with an opening on the front for bag handling (e. g.
sampling). The hood protects the attached bag from mechanical influences during
operation and reduces heat loss.
The hood is made from ABS and is also available in white. The hood should always be
used during operation.
Hood for Bag Holder 20:
Hood for Bag Holder 50:
The three-stage, Class A Pt100 resistance thermometer (EN60751) measures the
CultiBag RM temperature. The thermometer including the cable is run through the
opening into the Bag Holder 20 | 50 and connected to the Rocker 20 | 50 via a plug
connection. The head is secured in the opening via an O-ring. Connection to device:
Temp 1 and Temp 2, protection class: IP67
Make sure that the surface of the Pt100 sensor makes full contact with the
underside of the installed bag in order to ensure correct and safe system
operation.
Dimensions: 765 + 600 + 255 mm,
Weight: 2.5 kg
Dimensions: 1085 + 600 + 255 mm,
Weight: 3.7 kg.