The BIOSTAT
RM 20 | 50 basic can be used to aerate the cell culture with either
®
external compressed air or compressed room air.
A pressure sensor monitors the pressure and is equipped with an alarm signal
(see section
"7.2 Overpressure Control" on page 47).
t








A pressure sensor records the pressure value of the CultiBag RM (see section
"7.1 Schematic Representation of Aeration" on page 47).
t
If the pressure rises to > +30 mbar, the control value closes automatically until the
pressure decreases to +20 mbar. This protects the cell culture from overpressure and
protects the CultiBag RM from damage.





