4.4.6 Alarm Menu

4.4.7 Settings Menu

4.4.7.1 Setting the
Date and Time
34
Menu Operation
Alarm page:
This is accessed from the Main page via the "Alarm" button. You can view the last
alarms to occur by using the arrow keys to move through the list.
This screen is displayed automatically by the system when an alarm or message
occurs.
The alarm or message is confirmed by pressing the Home button.
For example, if an unfavorable status is detected for the reactor, i.e. overpressure in
the bag, the flow regulator is shut down and the Alarm page is displayed with the
current alarm along with its entry.
After leaving the Alarm page, the reason for the overpressure must be found and the
flow regulator restarted after correcting the problem.
Settings:
The "Settings" can be accessed from the Main page.
The date and time can be set here.
To do so, press the entry fields. This opens the keypad page for entry. Once all entries
have been made, press the "Set date and time" button to confirm.
This transfers the data to the CPU.