1.4 Rocker 20 | 50

1.4.1 Operation

y Hood:
White hood for Holder 20 and | or
White hood for Holder 50 and | or
Material: ABS
y Surface temperature sensor
Type Pt100/3 conductor | Class A
Protection class: IP67
Connection: Device | Temp1 | Temp2
Material: POM white, cable PFA, sensor surface chromium steel 1.4404
y Filter heater for Rocker 20 | 50
Electric resistance heating surface, 24 VDC | 6 W,
with automatic temperature control for approx. 48 °C
Caution!
Only use in combination with Rocker 20 | 50. Device connection FH1 | FH2.
y Tubing set for aeration
y Country-specific connection cable
y Quick reference guide
y CD-ROM with operating instructions and quick reference guide
CultiBags RM are not included in the scope of supply.
Bags must be ordered separately. When ordering, please specify the type
(Basic, Optical or Perfusion).
For more information, please contact your Sartorius Stedim Biotech
representative.
The rocker is a fully-automatic, PLC-controlled system which is operated via a touch
screen. The touch screen is on the front of the device and uses a hierarchical menu
structure with a backlight. The integrated functions include:
DS020L-R2WL
DS050L-R2WL
DZ-----R2PT
DZ-----R2FH
Design and Function