2.4 BIOSTAT
®
RM 20|50 basic

2.4.1 Connections, Interfaces

20
Delivery and Installation Instructions
Fig. 1: Schematic representation of the back of the BIOSTAT RM
Fig. 2: Schematic representation of the left side of the BIOSTAT RM
Fig. 3: Schematic representation of the right side of the BIOSTAT RM
Key:
SIG 1:
Reserve (LED connection or Loadcell)
D-link:
Optional Ethernet
SIG 2:
Alarm output, potential-free, max. 0.5 A
SIG 3:
Reserve (weight measurement)
D-LINK 2: DCU/ MFCS connection D-SUB
D-LINK 3: Optional Profibus DP D-SUB
AIR:
Compressed air connection max.1.5Bar | air pump intake
CO2:
Gas connection max. 1.5 (possibly O2)
T6.3A:
Main switch | Power connection,
100-240VDC broad band. Mains adapter, fuse 6.3A slow-acting
GAS1:
Gas mixture outlet to RM bag
TEMP 1:
PT100 temperature sensor connection
H 1:
Heating panel 48V | 3.15A
FH 1:
Filter heater 24V | 800mA
GAS 2:
Gas mixture outlet to RM bag
TEMP 2:
PT100 temperature sensor connection
H 2:
Heating panel 48V | 3.15A
FH 2:
Filter heater 24V | 800mA