Appendix
Alarm message
Possible causes
No CO
If the CO
2
(requires that the Gasflow1
controller is also active), this
alarm message is displayed
after 5 min when the CO
actual value is less than 0.6%
(target value min. 0.8%).
This does not deactivate the
CO
controller.
2
You can acknowledge this
alarm message via the "Home"
button.
Drive System Err
If no movement is detected in
the motor within a period of
15 s or an error is detected in
the motor amplifier (e.g.
overload), this alarm is
triggered.
This alarm is acknowledged by
turning the device off and on
(min. 5 s off). This message is
also displayed on the screen as
a ticker when there is an error.
Error Filter-H1
The filter heater is always in
operation when connected to
the system. However, it is only
monitored during active gas
aeration. In this case, it is
important that the filter
heater functions and this is
measured via the power
consumption. If filter heater 1
is disconnected from the
system (or defective) while the
Gasflow1 controller is active,
this alarm is triggered after a
few seconds. This does not
deactivate the controller. This
alarm message is only used 1x
per gas flow controller start.
You can acknowledge this
alarm message via the "Home"
button.
Measures
controller is active
Please check whether or not the
2
connection tubing to the device
has been crushed or kinked.
2
If the error continues to be
displayed even after a device
restart, you should contact a
service technician.
Make sure that the filter heater
is connected to the correct
output. If this error should
continue to occur, test the
function using another filter
heater.